It is used for analyzing the purity of oxyn in respirator machine of medical treatment and anesthesia
machine and oxyn concentrator and high-pressure oxyn house and the nursing case keeping infant.
Instrument sensor is electrochemistry diaphragmatic positron. It makes use of platinum negatron and silver and silver chloride as electrolyte.
Diaphragmatic material is polyethylene film which chooses the permeating oxyn .
Response time: 10 seconds to indicate 90% of oxyn concentration (at 20 ° C)
وصف
Specification:
Ran: 0-0-50% 0-1%
Accuracy: ± 0.10%
Use ambient temperature: -20~ 50 °C
Use: hand pinch suction type
Resolution: 0.01%
Power consumption: 4 mW
Power: 6F229V laminated battery, can work for 500 hours.
Affect the detection accuracy of interference gas: SO2, NO2, H2S 10% when the impact of the determination.
Dimensions: 23x18x7cm/9.06x7.09x2.76in
Packa Includes:
1 * Oxyn analyzer
1 * Oxyn sensor
1 * White filter hose
1 * Black suction ball
1* Portable carry bag