Precautions
1. Do not use dry boil
2. The instrument should be well grounded
3. When used for the first time, white peculiar smell come out, which is a normal phenomenon. Because the insulating material contains oil and other compounds during the production process, it can be placed in a ventilated place for a few minutes.
4. The temperature resistance of the motor and magnet of the blender is limited, so when you do the heating experiment, you cannot use it for heating alone. Please lower the motor speed to prevent the motor from being damaged by high temperature.
5. When there are wet hands, liquid overflow or long-term exposure to excessive humidity, induction electricity may be transmitted to the shell through the insulation layer, please ensure that it is grounded and pay attention to ventilation
6. Please keep the instrument clean if it is not used for a long time, and store it in a dry and non-corrosive gas
Specifications:
Input power/frequency: 110V/220V
Speed range: 50-2000 rpm, digital speed display
Motor power: 40W
Stirring capacity: 1000ml/2000ml/5000ml
Heating pot: φ160*80mm/φ190*90mm/φ240*150mm
Heating power: 300W/500W/750W
Heating temperature: water bath 100°C, oil bath 250°C
Temperature control accuracy: ±1°C
Temperature control method: intelligent digital display, controllable, internal and external sensor temperature measurement
Boiling: water -10-25 minutes
Working time: continuous
Package includes:
1 * magnetic stirrer